00122 VitalWrap Cervical Wrap |
 |
 |
The combination of heating, cooling, and compression have long been recognized in the medical field as a highly effective method for treating pain, inflamation, and swelling. The VitalWrap System is designed to deliver this therapy safely and conveniently, in a device that is easy to use and comfortable to wear. |
|
|
|
| |
|
|
 |
|
|
 |
|
|
| |
Features of the VitalWrap Hot & Cold Compression Therapy System
Set-up is a simple, three-step process:
Fill the portable, leak-proof reservoir with ice and water
Place the bladder portion of the wrap around the treatment area and secure it with the Velcro closure
Attach the wrap to the control unit using the tubing set Once the control unit is turned on, the cooled or warmed fluid is passed through the tubing set to the wrap. As it circulates into the wrap from the control unit, the fluid either applies heat to warm the treatment area or draws it away for deep cooling. The fluid then flows back into the control unit, where it is re-cooled or re-warmed before cycling back into the wrap. Compression is delivered through the wrap itself, and can continue even if the user chooses to disconnect the tubing set and control unit. The VitalWrap System is designed so that the user can apply thermal therapy for extended periods. This provides tremendous benefit to the injury site beyond the normal application periods of conventional therapies, such as heating pads and cold packs. The typical transition time between the warm and cold settings in the temperature range is about one minute. Specially tailored wraps are available for any area of the body. Please consult your healthcare professional prior to applying the system, particularly when using it for long periods of time.
Physical Size (approximately) 8"diameter x 13.5"H (20.3cm x 34.3cm)
Weight (dry) 5 lbs. (2.27 kg)
Fluid Tether 6 ft. (1.8m)
Thermal System Range 40°F to 105°F (4.4°C to 40.6°C)
Contrast therapy transition time 1 minute typical from extremes in temperature range
Operating period 2-6 hours typical continuous use
Thermal Cutout 47°C
Circulating System Reservoir Capacity 1.1 gallons (4.1 liters)
Reservoir Fluid:water
Flow Rate (through VitalWrap) 5.5 gph (21 lph)
Maximum Pressure 18 psi
Electrical System : Voltage 100, 120Vac, 50/60Hz, or 220-240Vac, 50Hz
Power 300W
Internally fuse System and pump
Leakage Current Under 300µA
Regulatory Classification: Class I Equipment Type BF
Regulatory Approvals:Compliance UL2601-1: Standard for Medical Equipment Part 1: General Requirements for Safety. CAN/CSA-C22.2 no. 601.1-M90: Medical Electrical Equipment Part 1: General Requirements for Medical Electrical Equipment. IEC 60601-1: Part 1: General Requirements for Safety. IEC 60601-2-35: Particular Requirements for the Safety of Blankets, Pads and Mattresses Intended for Heating in Medical Use |
|
|
|
 |
| |
 |
| |
|
|
|
|